Highly Precise (0,01%) Pressure Transmitters
{{details.product.PriceUnits}}
Please fill out this field...
Please enter a number...
Please enter at least 1...
Please enter at max 999999...

You purchased this item on {{details.product.OrderHistory[0].OrderDate | date : 'yyyy-MM-dd'}}

Digital Output of Transmitter:


This high precision of 0,01 %FS is available as an option (the standard Series 33 X has an accuracy of 0,05 %FS). These Series are based on the stable, floating piezoresisitive transducer and the newly developed XEMICS micro-processor with integrated 16 bit A/D converter. Temperature dependencies and non-linearities of the sensor are mathematically compensated.

Transmitter with Analog Output:


Integrated in the XEMICS processor is a D/A converter of 16 bit for analog signal outputs of 4…20 mA or 0…10 V. The output rate is 400 Hz. The accuracy is diminished by this converting process by 0,05 %FS. The digital output is available on all transmitters with analog output.

Programming

With the KELLER software READ30 and PROG 30, a RS485 converter (i.e. K102 or K107 from KELLER) and a PC, the pressure can be displayed, the units changed, a new gain or zero set. The analog output can be set to any range within the compensated range.

Accuracy and Precision

“Accuracy” is an absolute term, “Precision” a relative term. Dead weight testers are primary standards for pressure, where the pressure is defined by the primary values of mass, length and time. Highest class primary standards in national laboratories indicate the uncertainty of their pressure references with 70 to 90 ppM or close to 0,01%. Commercial dead weight testers as used in our facilities to calibrate the transmitters indicate an uncertainty or accuracy of 0,025%. Below these levels, KELLER use the expression “Precision” as the ability of a pressure transmitter to be at each pressure point within 0.01 %FS relative to these commercial standards. The transmitter’s full-scale output can be set up to match any standard of your choice by correcting the gain with the PROG30 software.

Pressure Ranges 0…0,3 to 0…1000 bar
Accuracy ±0,05 %FS
Total Error Band ±0,1 %FS @ -10…80 °C
Interfaces RS485, 4…20 mA, 0…10 V
Temperature Range -40…120 °C

 

 

 

 

 

 

 

 

For more information, click here.