Instrotech

LR-Cal LTC-FB-9018-60

LR-Cal LTC-FB-9018-60

Sku: LR-CAL-LTC-FB-9018-60

Portable temperature calibration microbath from -18°C to 140°C with ±0.15°C accuracy. Ideal for calibrating thermometers and temperature sensors

Product Information

The LR-Cal LTC-FB-9018-60 portable temperature calibration bath serves as both a temperature source and a reference instrument for testing and calibrating all types of temperature-measuring instruments. Various metal block inserts are also available.

  • Optimal for calibration of glass thermometers and laboratory thermometers as well as temperature sensors with not exactly cylindrical shape
  • Homogeneous temperature distribution due to the adjustable magnetic mixer
  • Simple operation with a colour touchscreen
  • Ramp function + thermostat tests
  • Computer-controlled calibrations via USB interface and optional software LR-Cal Aq2Sp2 for MS-Windows® computers
  • Option: changeable calibration bath insert for changing the calibration fluid without emptying and cleaning the reservoir
  • Option: dry block inserts to convert the device to a dry block temperature calibrator
  • Portable
  • Temperature range -18°C…+140°C
  • Accuracy ±0.15°C, stability ±0.02°C
  • Bath volume approx. 500 cm³
    (60 mm diameter, 170 mm depth)
  • Option: changeable calibration bath insert

LR-Cal LTC-FB-9018-60 specification

Description: Specification:
Temperature range: -18…+140°C
Display + operation: colour touchscreen
Display accuracy: ±0.15°C *
Measuring units: °C / °F / K
Displey resolution: 0.1° / 0.01° / 0.00^°
Mean heating time:
(incl. stabilization)
approx. 50 min.
from ambient temperature to +140°C
Mean cooling time:
(incl. stabilization)
approx. 50 min.
from +140°C to ambient temperature
Stability: ±0.02°C at 0°C **
Axial uniformity: ±0.05°C
Radial uniformity: ±0.07°C at +140°C ***
Hole diameter: 60 mm
Hole depth: 170 mm
Insert material: anticorodal
Switch test; voltage: Open/Closed; 4.5 VDC
Adjustable ramp function: from 0.1°C/min
Computer interface: USB type B
Automatic calibration: at 5 temperature points
External sensors/probes: Optional with Order-Code LTC-MP-3I (to be ordered together with instrument)
Operation voltage: 230 VAC (optional 115 VAC) 50/60 Hz
Electric powert: 300 VA
Dimensions: 160 x 360 x 350 mm
Weight incl. standard accessories: 11.0 kg

All values have been measured at ambient temperature of +20°C and with a sensor with diamter 6 mm.
The achievement of stabilization is confirmec by a displayed root symbol and an audial beep.
* = Temperature deviation between display and reference probe.
** = Max. temperature difference at stable temperature over 30 min.
*** = Measured from the button up to 40 mm.