Option time and date stamp

CThis option is only available for the models 5004 and 6004. This option allows the user to time and date stamp the display value either by momentarily pressing the ‘menu’ key, or by transmitting a byte (any ASCII character) to the unit. This option is very useful to interface to serial printers.

Data Protocol Format
Protocol format is: #_HH:MM:SS_DD/MM/YY_S_ _DDDDDD_CR/LF

  • Decimal point is in the same position as the display decimal point.
  • See option 3002/3013 for more information on the Asciibus Protocol.

Protocol format is

  • _ = Space
  • S = Sign
  • # = indicates start of message
  • D = data for 6 numerals.
  • CR = ASCII carriage return.
  • LF = ASCII line feed

NOTE:
- The time is only available in 24hr mode.
- This option comes standard with options 3013 (Rs232 Communications Option).