{{image.AltTag}} {{image.AltTag}}
Default Image Default Image

{{details.product.ModelName}}

Sku: {{details.product.ModelNumber}}

Infrared thermometer optris CTlaser MT for temperature measurement through flames
Please select an item in the list...
{{details.product.PriceUnits}}
Please fill out this field...
Please enter a number...
Please enter at least 1...
Please enter at max 999999...

You purchased this item on {{details.product.OrderHistory[0].OrderDate | date : 'yyyy-MM-dd'}}

The infrared thermometer optris CTlaser MT offers a special measuring wavelength for precise temperature measurement through flames of 200°C to 1,650°C. It is thus perfectly suited for surveillance of work pieces in ovens, for measurements in chemical reactors and for testing of refractory linings in kilns.

The stainless steel measuring head with its highly precise double laser visor allows for exact marking of measurement spots in any distance at any time. The temperature of even the smallest objects with a size of only 1.6 mm can be easily measured with the IR thermometer.

  • Temperature range:
    200°C to 1,450°C
  • Spectral range: 3.9 µm
  • Response time: 10 ms

Applications

The two-part IR thermometer CTlaser MT succeeds where most other pyrometers fail: in measuring temperatures through glass and flames.

This feature makes it especially interesting for industrial high-temperature applications with temperatures up to 1,650°C and it is thus employed in e.g. surveillance of metallic work pieces in ovens, measurements in chemical reactors, or temperature controls of refractory linings in kilns.

Software

The temperature analysis software optris Compact Connect has been developed specifically for the extensive documentation and analysis of temperature measurement data of our stationary infrared thermometers. It is license-free and included in every pyrometer shipment.   

Measurement Specifications

  • Temperature range (scalable via programming keys or software): 200°C ... 1,650°C
  • Spectral range: 3.9 µm
  • Optical resolution (90% energy): 45:1
  • Selectable focus:

    CF1 1.6 mm @70 mm
    CF2 3.4 mm @ 150 mm
    CF3 4.5 mm @ 200 mm
    CF4 10 mm @ 450 mm
    SF 27 mm @ 1200 mm

  • System accuracy1) (at Tamb= 23 ± 5°C): ± 1%2)
  • Repeatability (at Tamb = 23 ± 5°C): ±0.5% or ±0.5°C2)
  • Temperature resolution (digital): 0.1 K
  • Response time (90% signal)3): 10 ms
  • Emissivity / Gain (adjustable via programming keys or software): 0.100 - 1.100
  • Transmissivity / Gain (adjustable via programming keys or software): 0.100 - 1.000
  • Signal processing (parameter adjustable via programming keys or software, respectively): 
    peak hold, valley hold, average, extended hold function with threshold and hysteresis

General Specifications

  • Environmental rating: IP 65 (NEMA-4)
  • Ambient temperature: 
    sensing head: -20°C ... 85°C (50°C with laser ON)
    electronics: -20°C ... 85°C
  • Storage temperature:
    sensing head: -40°C ... 85°C
    electronics: -40°C ... 85°C
  • Relative humidity: 10 - 95%, non condensing
  • Vibration: IEC 60068-2-6 (sinus shaped), IEC 60068-2-64 (broadband noise)
  • Shock: IEC 60068-2-27 (25G and 50G)
  • Weight:
    sensing head: 600 g
    electronics: 420 g

Electrical Specifications

  • Outputs / analog:
    channel 1: 0/4 - 20 mA, 0 - 5/10 V, thermocouple J, K
    channel 2: sensing head temperature (-20 - 85°C as 0 - 5 V or 0 - 10 V), alarmoutput
  • Output / alarm: 24 V / 50 mA (open collector)
  • Optional: relay: 2 x 60 V DC/42 V ACeff; 0.4 A; optically isolated
  • Outputs/ digital (optional): USB, RS232, RS485, Modbus RTU, Profibus DP, Ethernet
  • Output impedances:

    mA max. 500 Ω (with 5 - 36 V DC)
    mV min. 100 kΩ load impedance
    thermocouple 20 Ω

  • Inputs: programmable functional inputs for external emissivity adjustment, ambient temperature compensation, trigger (reset of hold functions)
  • Cable length: 3 m (standard), 8 m, 15 m
  • Power supply: 8-36 V DC
  • Current draw: max. 160 mA
  • Laser 635 nm: 1mW, ON/OFF via electronic box or software

Versions

Part number: temperature range / spectral range / optics / response time

  • OPTCTLMT: 200-1450°C / 3.9 μm / 45:1 / 10 ms
  • OPTCTLMTH: 400-1650°C / 3.9 μm / 45:1 / 10 ms